Fresenius Liberty Select Cycler
System, Peritoneal, Automatic Delivery. 510(k) K171652, from Fresenius Medical Care Renal Therapies Group, LLC (Waltham, MA). Product code FKX, class 2.
As of Nov 27, 2026, K171652 is an FDA 510(k) held by Fresenius Medical Care Renal Therapies Group, LLC, substantially equivalent on Sep 15, 2017 after 102 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 102 days
- Recalls: 2
- Cited as a predicate by: 1 device
The Fresenius Liberty Select Cycler is indicated for acute and chronic peritoneal dialysis.
- Fresenius Liberty CyclerK123630 · Fresenius Medical Care North America, Design Cente · Sep 9, 2013
- Liberty Pdx CyclerK141145 · Fresenius Medical Care Renal Therapies Group, LLC · Jul 30, 2014
- Fresenius Liberty Select CyclerK181108 · Fresenius Medical Care Renal Therapies Group, LLC · May 24, 2018
| Sep 12, 2020 | Z-0439-2021The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not… | Completed |
| Jan 24, 2018 | Z-1365-2018The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result… | Terminated |
- Archimedes Pro Automated Peritoneal Dialysis Cycler (3014-00000-1); Archimedes Standard Automated Peritoneal Dialysis…K250523 · Simergent, LLC · Oct 2, 2025
- Byonyks X-1 APD Cycler; Byonyks Automated PD Set DS-1K243371 · Byonyks Pvt, Ltd. · May 16, 2025
- Digital Dialysis Clinic - Intelligent Dialysis Assistant (IDA) with Medical Supervision Portal and IDA Smartphone AppK220935 · Iren-Medical, Ltd. · Dec 29, 2022
- Fresenius Liberty Select CyclerK222318 · Fresenius Medical Care Renal Therapies Group, LLC · Oct 31, 2022
- Lilliput APD SystemK212522 · Fresenius Medical Care Renal Therapies Group, LLC · Apr 15, 2022
- CloudCath Peritoneal Dialysis Drain Set Monitoring SystemK212658 · Cloudcath · Feb 9, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 5, 2017
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5630
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files