System, Planning, Radiation Therapy Treatment. 510(k) K171536, from RaySearch Laboratories AB (PUBL) (Stockholm, SE). Product code MUJ, class 2.

As of Nov 20, 2026, K171536 is an FDA 510(k) held by RaySearch Laboratories AB (PUBL), substantially equivalent on Jun 27, 2017 after 33 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 33 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • RayStation 6K170355 · RaySearch Laboratories AB (PUBL) · Mar 2, 2017

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • RayStation 7K180379 · RaySearch Laboratories AB (PUBL) · Apr 25, 2018

Recalls

Recalls FDA ties to K171536, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 8, 2025Z-2494-2025DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.Open
Jan 16, 2025Z-1192-2025Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…Open
Mar 28, 2018Z-2290-2018The dose calculation accuracy may in some situations be less than expected. The user must be aware in order to avoid incorrect dose calculations during treatment planning.Terminated
Jan 26, 2018Z-1275-2018Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient.Terminated
Nov 20, 2017Z-1930-2019When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrongTerminated
Jul 17, 2017Z-0380-2018For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
May 25, 2017
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files