ShockPulse-SE Lithotripsy System
Lithotriptor, Ultrasonic. 510(k) K171024, from Cybersonics, Inc. (Erie, PA). Product code FEO, class 2.
As of Nov 22, 2026, K171024 is an FDA 510(k) held by Cybersonics, Inc., substantially equivalent on Aug 21, 2017 after 138 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 138 days
- Recalls: 6
ShockPulse-SE Lithotripsy System is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter and bladder.
- Shock Pulse-SE Lithotripsy System CYBERWAND IIK142428 · Cybersonics, Inc. · Nov 13, 2014
| Jan 8, 2026 | Z-1494-2026Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards… | Open |
| Nov 14, 2025 | Z-1478-2026, Z-1479-2026, Z-1477-2026Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these… | Open |
| Oct 8, 2025 | Z-0452-2026The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in… | Open |
| Jul 7, 2022 | Z-1514-2022Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards… | Open |
- Swiss LthoClast TrilogyK230893 · E.M.S Electro Medical Systems S.A · Oct 23, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 5, 2017
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.4480
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files