Arca
Skip to content
Tools
Tools

CyberKnife Treatment Delivery System

Accelerator, Linear, Medical. 510(k) K170788, from Accuray Incorportaed (Sunnyvale, CA). Product code IYE, class 2.

As of Nov 20, 2026, K170788 is an FDA 510(k) held by Accuray Incorportaed, substantially equivalent on Apr 11, 2017 after 26 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 26 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CyberKnife® Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K170788, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 16, 2026Z-2640-2026, Z-2644-2026In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room…Open
Dec 23, 2024Z-1023-2025Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.Completed
Feb 26, 2021Z-1272-2021The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Elsewhere

Record

Type
Traditional
Received
Mar 16, 2017
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files