Accelerator, Linear, Medical. 510(k) K170751, from Viewray, Incorporated (Oakwood Village, OH). Product code IYE, class 2.

As of Nov 20, 2026, K170751 is an FDA 510(k) held by Viewray, Incorporated, substantially equivalent on Jun 7, 2017 after 86 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K170751, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 20, 2020Z-0482-2021There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.Terminated
Nov 18, 2019Z-1803-2020Registering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy and structures. The MRIdian…Terminated
Mar 18, 2019Z-1251-2019A discrepancy between optimization and planning forward dose calculation between adaptive optimizations and AQA dose calculations can occur.Terminated
Mar 3, 2019Z-0676-2020Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Mar 13, 2017
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files