Arca
Skip to content
Tools
Tools

Invacare® TDX® SP2 Power Wheelchair and 1 more

Wheelchair, Powered. 510(k) K170507, from Invacare Corporation (Elyria, OH). Product code ITI, class 2.

As of Nov 18, 2026, K170507 is an FDA 510(k) held by Invacare Corporation, substantially equivalent on Jul 12, 2017 after 141 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 141 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K170507, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 30, 2022Z-1130-2022The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The…Open
Nov 12, 2021Z-0445-2022Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile,…Open

Other ITI clearances

The newest 510(k)s for the same product code; their median review was 188 days; see every ITI device

Record

Type
Traditional
Received
Feb 21, 2017
Review panel
Physical Medicine
Regulation
21 CFR 890.3860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files