Flextra Steerable Introducer
Introducer, Catheter. 510(k) K170373, from Rhythm Xience, Inc. (Rxi) (Eden Prairie, MN). Product code DYB, class 2.
As of Nov 21, 2026, K170373 is an FDA 510(k) held by Rhythm Xience, Inc. (Rxi), substantially equivalent on Feb 27, 2017 after 20 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 20 days
- Recalls: 4
- Cited as a predicate by: 2 devices
The Flextra™ Steerable Introducer is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
- Agilis Nxt Steerable IntroducerK061363 · St Jude Medical · Jul 12, 2006
- AcQCross Qx Integrated Transseptal Dilator/NeedleK210685 · Acutus Medical, Inc. · Apr 6, 2021
- AcQGuide Flex with AcQCross QX, AcQGuide Mini with AcQCross QXK193509 · Acutus Medical, Inc. · Jan 17, 2020
| Started | Recall | Status |
|---|---|---|
| Feb 25, 2020 | Z-1771-2020, Z-1770-2020, Z-1769-2020, Z-1768-2020It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism. | Terminated |
- Giranel Guiding Sheath SetK262674 · Kossel Medtech (Suzhou) Co., Ltd. · Oct 15, 2026
- NAMIC Dilator and SheathK253526 · Medline Industries, LP · Jul 29, 2026
- Micropuncture Kit / Micropuncture Access Kit / Micropuncture Introducer Set/ Microintroducer Kit / Microintroducer Kit…K261529 · Hartman Medical Group, LLC · Jul 6, 2026
- SimplGuideK261242 · Pendracare · Jun 25, 2026
- Health Line Tearaway IntroducerK261950 · Health Line International Corporation · Jun 22, 2026
- FieldFlex Steerable SheathK261492 · CenterPoint Systems, LLC · Jun 5, 2026
- Echo Large Bore Introducer SheathK260606 · Echo Medical, LLC · Apr 29, 2026
- Introducer Sheath SetK254248 · Cardiocycle Medical(Suzhou) Co., Ltd. · Apr 23, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Feb 7, 2017
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1340
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files