RA-308 Excimer Laser System, Excimer Laser Catheter

Catheter For Crossing Total Occlusions. 510(k) K170349, from Ra Medical Systems,Inc (Carlsbad, CA). Product code PDU, class 2.

As of Nov 19, 2026, K170349 is an FDA 510(k) held by Ra Medical Systems,Inc, substantially equivalent on May 24, 2017 after 110 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 110 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The DABRA Laser System is indicated for crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K170349, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 16, 2020Z-0569-2021Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue…Terminated
Jul 24, 2020Z-0050-2021Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.Terminated
Jan 23, 2020Z-2497-2020A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user…Terminated
Sep 19, 2019Z-2006-2020The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.Terminated
Feb 15, 2018Z-2218-2019Lasers/Catheters did not calibrate during set-up prior to use.Terminated

Other PDU clearances

The newest 510(k)s for the same product code; their median review was 96 days; see every PDU device

Record

Type
Traditional
Received
Feb 3, 2017
Review panel
Cardiovascular
Regulation
21 CFR 870.1250

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files