Arca
Skip to content

Neuronaute

Full-Montage Standard Electroencephalograph. 510(k) K170138, from Bioserenity (Paris, FR). Product code GWQ, class 2.

As of Nov 17, 2026, K170138 is an FDA 510(k) held by Bioserenity, substantially equivalent on Dec 26, 2017 after 343 days, with no recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 26, 2017
  • Review time: 343 days
  • Recalls: 0
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Neuronaute is a medical, wireless and mobile equipment, which allows acquisition, record, storage, transmission and displaying of electroencephalogram (EEG) from adult patients. It can be used with patients in health care facility (data acquisition and reporting) or clinical research environment. The Neuronaute requires operation by a trained healthcare professional.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • Dreem 3SK223539 · Beacon Biosignals, Inc. · Aug 18, 2023

Other GWQ Clearances

The newest 510(k)s for the same product code; their median review was 182 days; see every GWQ device

Record

Type
Traditional
Received
Jan 17, 2017
Review panel
Neurology
Regulation
21 CFR 882.1400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.