Hoffmann LRF System

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K163656, from Stryker GmbH (Mahwah, NJ). Product code KTT, class 2.

As of Nov 17, 2026, K163656 is an FDA 510(k) held by Stryker GmbH, substantially equivalent on Apr 14, 2017 after 112 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 14, 2017
  • Review time: 112 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Stryker Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: • Open and Closed Fractures • Post-traumatic joint contracture which has resulted in loss of range of motion • Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction • Pseudoarthrosis or non-union of long…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K163656, newest first; each opens FDA's recall record
Aug 28, 2018Z-0290-2019The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin.…Terminated
Aug 28, 2018Z-0289-2019The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin.…Terminated
Aug 28, 2018Z-0291-2019The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin.…Terminated
Aug 28, 2018Z-0288-2019The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin.…Terminated
Jul 26, 2018Z-3240-2018When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction…Terminated

Other KTT Clearances

The newest 510(k)s for the same product code; their median review was 84 days; see every KTT device

Record

Type
Traditional
Received
Dec 23, 2016
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.