IMPIX 3D Print Cages
Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K163595, from Medicrea International SA (Rillieux-La-Pape, FR). Product code MAX, class 2.
As of Nov 17, 2026, K163595 is an FDA 510(k) held by Medicrea International SA, substantially equivalent on Nov 13, 2017 after 327 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 13, 2017
- Review time: 327 days
- Recalls: 2
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIMPIX 3D Print cages are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Cascadia Interbody SystemK150481 · K2m, Inc. · Aug 20, 2015
- Impix SpacerK072226 · Medicrea Technologies · Dec 10, 2007
- Capstone Spinal SystemK121760 · Medtronic Sofamor Danek USA, Inc. · Aug 29, 2012
- Impix Alif Lumbar Ibf, Impix Alif-D Lumbar Ibf, Impix Tlif Lumbar IbfK083798 · Medicrea Technologies · May 19, 2010
- ALIF Interfixated SystemK151214 · Nu Vasive, Incorporated · Jun 25, 2015
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- IB3D™ PL Spinal SystemK241164 · Medicrea International S.A.S. (Medtronic) · Sep 6, 2024
- IB3D AlifK191134 · Medicrea International SA · Oct 30, 2019
- UNiD Patient specific 3D printed TLIF cageK190092 · Medicrea International SA · May 8, 2019
- UNiD Patient-matched PLIF cageK182158 · Medicrea International SA · Jul 15, 2019
Recalls
Recalls FDA ties to K163595, newest first; each opens FDA's recall record| Jul 12, 2023 | Z-2452-2023There is a potential for a product packaging non-conformity issue which presents as a pinhole in either the inner or outer pouch. | Open, Classified |
| Oct 2, 2019 | Z-1156-2020Inner draw rod, which threads into the cage may potentially be separated from the proximal knob of the inserter | Terminated |
Other MAX Clearances
The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device- Symmetrix PX Spinal SystemK262572 · Medtronic Sofamor Danek USA, Inc. · Oct 29, 2026
- Calibrate MAX System; Calibrate NanoTec MAX System; Calibrate CCX Interbody System;…K262646 · Alphatec Spine, Inc. · Oct 21, 2026
- Scala® TlK261049 · Spineart SA · Aug 20, 2026
- TruLift Expandable SystemK261430 · Life Spine, Inc. · Jul 28, 2026
- Catalyft PL Expandable Interbody SystemK253011 · Medtronic Sofamor Danek USA, Inc. · Jul 17, 2026
- Symmetrix L Spinal SystemK261020 · Medtronic Sofamor Danek USA, Inc. · Jul 14, 2026
- Admiral ACP System; Explorer TO System; Laminoplasty System; Manta Ray TDF Spacer;…K253420 · SeaSpine Orthopedics Corporation · Jul 1, 2026
- Fule Interbody Fusion Cage SystemK253151 · Beijing Fule Science & Technology Development Co., Ltd. · Jun 15, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 21, 2016
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3080
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files