EXPRT™ Revision Hip System - EXPRT™ Hip Distal Stem, EXPRT™ Hip Standard Offset Proximal Body Implant, EXPRT™ Hip Lateral Offset Proximal Body Implant, EXPRT™ Hip Capture Bolt
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer. 510(k) K163497, from Encore Medical L.P. (Austin, TX). Product code KWZ, class 2.
As of Nov 18, 2026, K163497 is an FDA 510(k) held by Encore Medical L.P., substantially equivalent on Mar 2, 2017 after 79 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Mar 2, 2017
- Review time: 79 days
- Recalls: 1
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFEXPRRTw Revision Hip System - EXPRRTw Hip Distal Stem, EXPRRTw Hip Standard Offset Proximal Body Implant, EXPRRTw Hip Device Name 510(k) Number (if(xrown)) K163497 *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, see existing data…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- EXPRT Revision Hip SystemK161610 · Encore Medical L.P. · Nov 9, 2016
Recalls
Recalls FDA ties to K163497, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Sep 21, 2018 | Z-1123-2019Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision… | Terminated |
Other KWZ Clearances
The newest 510(k)s for the same product code; see every KWZ device- Mpact Constrained LinerK241461 · Medacta International S.A. · Jul 31, 2025
- PINNACLETM Constrained Acetabular LinersK240639 · Depuy Ireland UC · Apr 4, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 13, 2016
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3310
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files