OrthoSolutions 'System26' Bone Screws
Screw, Fixation, Bone. 510(k) K163489, from Ortho Solutions UK Limited (Maldon, GB). Product code HWC, class 2.
As of Nov 21, 2026, K163489 is an FDA 510(k) held by Ortho Solutions UK Limited, substantially equivalent on Mar 8, 2017 after 86 days, with 1 recall tied to it in FDA's recall database.
- Review time: 86 days
- Recalls: 1
- Cited as a predicate by: 2 devices
The OrthoSolutions ‘System26’ cannulated screws (headed and headless compression) and washers are indicated for use over a guide pin or wire for aligned bone fracture repair and arthrodesis, osteotomy, joint fusion and bone fragment fixation appropriate with the size of the screw.
- Intraosseous Fusion Device SystemK231649 · Auxano Medical, LLC · Sep 29, 2023
- BRM TOOL ScrewsK203773 · Brm Extremities Srl · Jan 14, 2022
| Started | Recall | Status |
|---|---|---|
| Oct 10, 2017 | Z-0967-2018Following the identification of corrosion residue found inside the cannulated devices in Forefoot and Midfoot trays Orhto Solutions Group has taken the decision to recall ALL x106 System 26 Forefoot and Midfoot Implants… | Terminated |
- DuoCore Non-Compressive IM NailK263241 · GM Dos Reis Industria e Comercio Ltda. · Nov 2, 2026
- Ntegrate Threaded Nail SystemK261961 · Nvision Biomedical Technologies · Oct 14, 2026
- OsteoCentric Technologies Cannulated and Soild Fasteners with NutsK263116 · OsteoCentric Technologies · Oct 2, 2026
- Arthrex Beaming SystemK262975 · Arthrex, Inc. · Sep 25, 2026
- DynaNail Helix® Fixation SystemK260995 · MedShape, Inc. · Sep 23, 2026
- F.A.S.T ScrewK254280 · Steps Ortho · Sep 21, 2026
- ActivaScrew™ Cannulated COMK261057 · Bioretec, Ltd. · Sep 4, 2026
- OsteoCoil™ GT Active Compression SystemK260106 · Flower Orthopedics Corporation DBA Conventus Flower Orthoped · Aug 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 12, 2016
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files