Arca
Skip to content
Tools
Tools

Confirm Rx Insertable Cardiac Monitor (ICM) System

Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection). 510(k) K163407, from St. Jude Medical, Inc. (Sylmar, CA). Product code MXC, class 2.

As of Nov 20, 2026, K163407 is an FDA 510(k) held by St. Jude Medical, Inc., substantially equivalent on Sep 29, 2017 after 298 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 298 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Confirm Rxâ„¢ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K163407, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 19, 2018Z-1220-2019The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.Terminated
Apr 16, 2018Z-2335-2018, Z-2334-2018The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac…Terminated
Apr 13, 2018Z-0808-2020Devices inappropriately displayed a low battery indicator determined to be due to memory corruption.Terminated

Other MXC clearances

The newest 510(k)s for the same product code; see every MXC device

Record

Type
Traditional
Received
Dec 5, 2016
Review panel
Cardiovascular
Regulation
21 CFR 870.2800

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.