Arca
Skip to content
Tools
Tools

HAMILTON-H900, HAMILTON-BC8010, HAMILTON-BC4010

Humidifier, Respiratory Gas, (Direct Patient Interface). 510(k) K163283, from Hamilton Medical AG (Bonaduz, CH). Product code BTT, class 2.

As of Nov 18, 2026, K163283 is an FDA 510(k) held by Hamilton Medical AG, substantially equivalent on Sep 20, 2017 after 303 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Sep 20, 2017
  • Review time: 303 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Intended Use HAMILTON-H900 The HAMILTON-H900 humidifier is intended to add moisture and to warm breathing gases during invasive and non-invasive mechanical ventilation. The intended area of use is the intensive care ward or the recovery room.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K163283, newest first; each opens FDA's recall record
StartedReasonStatus
Jan 5, 2021Z-1385-2022When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm…Completed

Other BTT Clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every BTT device

Record

Type
Traditional
Received
Nov 21, 2016
Review panel
Anesthesiology
Regulation
21 CFR 868.5450

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files