Arca
Skip to content
Tools
Tools

Neodent Implant System - GM Line

Implant, Endosseous, Root-Form. 510(k) K163194, from Jjgc Industria E Comercio DE Materiais Dentarios S.A. (Curitiba, BR). Product code DZE, class 2.

As of Nov 18, 2026, K163194 is an FDA 510(k) held by Jjgc Industria E Comercio DE Materiais Dentarios S.A., substantially equivalent on Jul 14, 2017 after 241 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 14, 2017
  • Review time: 241 days
  • Recalls: 6
  • Cited as a predicate by: 23 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Indications for Use for GM implants and conventional abutments: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

23 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K163194, newest first; each opens FDA's recall record
Aug 12, 2026Z-3173-2026There is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside.Open, Classified
Jul 13, 2026Z-2929-2026There may be a mismatch between the implant contained inside the packaging and the information described on the outer carton label. If the the incorrect implant is installed, the issue could cause bone compression or…Open, Classified
May 1, 2025Z-1862-2025It is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead.Open, Classified
May 1, 2025Z-1863-2025It is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead.Open, Classified
Oct 11, 2021Z-0203-20223.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is…Terminated
Oct 2, 2018Z-0455-2019Laser engraved label does not match with the item in the packageTerminated

Other DZE Clearances

The newest 510(k)s for the same product code; their median review was 257 days; see every DZE device

Record

Type
Traditional
Received
Nov 15, 2016
Review panel
Dental
Regulation
21 CFR 872.3640

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files