VOCSN Unified Respiratory System
Ventilator, Continuous, Facility Use. 510(k) K162877, from Ventec Life Systems, Inc. (Bothell, WA). Product code CBK, class 2.
As of Nov 20, 2026, K162877 is an FDA 510(k) held by Ventec Life Systems, Inc., substantially equivalent on Apr 7, 2017 after 175 days, with 16 recalls tied to it in FDA's recall database.
- Review time: 175 days
- Recalls: 16
- Cited as a predicate by: 1 device
The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non-invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg.
- Trilogy Series Ventilator With OximetryK111610 · Respironics, Inc. · Oct 14, 2011
- Palmtop Ventilator, Models PTV-8, PTV-10K070594 · Pulmonetic Systems, Inc. · Aug 1, 2007
- Ltv 1200 VentilatorK060647 · Pulmonetic Systems, Inc. · May 25, 2006
- Omni 3 Oxygen SystemK120785 · Caire, Inc. · Oct 25, 2012
- Coughassist T70 DeviceK121955 · Respironics, Inc. · Dec 14, 2012
- Easy Go VacK140179 · Precision Medical, Inc. · Apr 25, 2014
- Trilogy EvoK181166 · Respironics, Inc. · Jul 18, 2019
| Sep 17, 2026 | Z-0129-2027, Z-0123-2027, Z-0128-2027, Z-0130-2027, Z-0125-2027, Z-0126-2027, Z-0127-2027, Z-0122-2027, Z-0124-2027, Z-0131-2027, Z-0132-2027Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. | Open |
| May 8, 2026 | Z-2525-2026Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of… | Open |
| Mar 25, 2026 | Z-1942-2026Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could… | Open |
| Feb 3, 2025 | Z-1196-2025Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on… | Open |
| Jan 9, 2024 | Z-1070-2024The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions… | Open |
| May 10, 2021 | Z-1902-2021Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy. | Open |
- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 14, 2016
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files