iTotal® Hip Replacement System
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented. 510(k) K162719, from Conformis, Inc. (Bedford, MA). Product code LPH, class 2.
As of Nov 17, 2026, K162719 is an FDA 510(k) held by Conformis, Inc., substantially equivalent on Jun 14, 2017 after 258 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 14, 2017
- Review time: 258 days
- Recalls: 4
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe iTotal® Hip Replacement System is designed from a patient’s pre-operative CT scan which must include certain necessary anatomic landmarks that are clearly identifiable. Total hip replacement using the iTotal® Hip Replacement System is indicated for use in skeletally mature individuals undergoing total hip replacement due to: • A severely painful and/or disabled joint from osteoarthritis…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Pipeline Total Hip SystemK112802 · Pipeline Orthopedics · Mar 9, 2012
- Corail Amt Hip ProsthesisK042992 · DePuy Orthopaedics, Inc. · Feb 11, 2005
- Signature Planner,Signature GuidesK111863 · Materialise NV · Jun 15, 2012
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Cordera Hip SystemK202484 · Conformis, Inc. · Sep 24, 2020
- Conformis Hip SystemK192198 · Conformis, Inc. · Nov 6, 2019
- Conformis BeneFIT Hip SystemK190904 · Conformis, Inc. · Aug 28, 2019
Recalls
Recalls FDA ties to K162719, newest first; each opens FDA's recall record| Aug 22, 2019 | Z-0063-2020Two femoral stems, sizes 13 and 15, were erroneously switched and packed in each other's packaging. | Terminated |
| Aug 22, 2019 | Z-0064-2020Two femoral stems, sizes 13 and 15, were erroneously switched and packed in each other's packaging. | Terminated |
| Mar 29, 2019 | Z-1775-2019Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants. | Terminated |
| Oct 16, 2018 | Z-0731-2019The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm. | Terminated |
Other LPH Clearances
The newest 510(k)s for the same product code; their median review was 282 days; see every LPH device- G7® Acetabular SystemK261453 · Zimmer Biomet · Sep 23, 2026
- Mpact 3D Metal FoamK253630 · Medacta International S.A. · Aug 28, 2026
- TRIVANCE Hip StemK254043 · Globus Medical, Inc. · Aug 4, 2026
- Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained Acetabular Insert; Trident®…K261624 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 10, 2026
- iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi CoatingK243192 · Nextstep Arthopedix, LLC · Jun 5, 2026
- Taperloc® Complete Hip StemK251270 · Biomet, Inc. · May 19, 2026
- G7® Revision Acetabular SystemK252623 · Zimmer, Inc. · Feb 4, 2026
- Mpact 3D Metal Implants Extension DMLS TechnologyK251043 · Medacta International S.A. · Jan 28, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 29, 2016
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3358
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files