VERIFY Assert Self-Contained Biological Indicator
Biological Sterilization Process Indicator With Recombinant-Dna Plasmid. 510(k) K162701, from STERIS Corporation (Mentor, OH). Product code OWP, class 2.
As of Nov 21, 2026, K162701 is an FDA 510(k) held by STERIS Corporation, substantially equivalent on Mar 24, 2017 after 177 days, with no recalls tied to it in FDA's recall database.
- Review time: 177 days
- Recalls: 0
- Cited as a predicate by: 5 devices
The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing and product testing of the following steam sterilization processes. Cycle Type Temperature Time Dynamic Air Removal 270°F (132°C) 4 minutes Dynamic Air Removal 275°F (135°C) 3 minutes Gravity 250°F (121°C) 30 minutes Gravity 270°F (132°C) 15 minutes When used in conjunction with the…
- K102469K102469
- Verify Cronos Self-Contained Biological IndicatorDEN110006 · STERIS Corporation · Jul 12, 2013
- VERIFY Assert Self-Contained Biological IndicatorK200126 · STERIS Corporation · May 22, 2020
- VERIFY Assert Self-Contained Biological IndicatorK181442 · STERIS Corporation · Jun 13, 2018
- VERIFY Incubator for Assert VH2O2 Self-Contained Biological IndicatorK171587 · STERIS Corporation · Aug 22, 2017
- VERIFY ASSERT STEAM Process Challenge Device for Gravity CyclesK170070 · STERIS Corporation · Apr 28, 2017
- VERIFY Incubator for Assert Self Contained Biological IndicatorsK163587 · STERIS Corporation · Mar 24, 2017
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 28, 2016
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.2805
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files