Arca
Skip to content
Tools
Tools

Allura Xper R9

Interventional Fluoroscopic X-Ray System. 510(k) K162148, from Philips Medical Systems Nethrlands BV (Best Noord-Brabant, NL). Product code OWB, class 2.

As of Nov 18, 2026, K162148 is an FDA 510(k) held by Philips Medical Systems Nethrlands BV, substantially equivalent on Nov 23, 2016 after 114 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 114 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Allura Xper R9 series (within the limits of the used Operating Room table) are intended for use to perform: • Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K162148, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 13, 2021Z-2476-2021Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar…Terminated
Dec 23, 2019Z-1145-2020Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.Terminated
Aug 18, 2017Z-0255-2018When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Abbreviated
Received
Aug 1, 2016
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files