Abutment, Implant, Dental, Endosseous. 510(k) K161435, from Nobel Biocare AB (Goteborg, SE). Product code NHA, class 2.

As of Nov 17, 2026, K161435 is an FDA 510(k) held by Nobel Biocare AB, substantially equivalent on Nov 22, 2016 after 182 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Nov 22, 2016
  • Review time: 182 days
  • Recalls: 1
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Temporary Snap Abutment is intended to be used to fabricate and support provisional restorations that aid in creating an esthetic emergence through the gingiva during the healing period and prior to final restoration. The Temporary Snap Abutment can be used for cement retained or screw-retained provisional restorations. The abutments can be used for single-unit and multi-unit restorations.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K161435, newest first; each opens FDA's recall record
StartedReasonStatus
Aug 9, 2019Z-1857-2020Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct…Terminated

Other NHA Clearances

The newest 510(k)s for the same product code; their median review was 169 days; see every NHA device

Record

Type
Traditional
Received
May 24, 2016
Review panel
Dental
Regulation
21 CFR 872.3630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.