Unit, Electrosurgical, Endoscopic (With Or Without Accessories). 510(k) K160417, from Nxthera, Inc. (Maple Grove, MN). Product code KNS, class 2.

As of Nov 17, 2026, K160417 is an FDA 510(k) held by Nxthera, Inc., substantially equivalent on Mar 17, 2016 after 30 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Mar 17, 2016
  • Review time: 30 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm 3 and ≤ 80cm 3. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K160417, newest first; each opens FDA's recall record
StartedReasonStatus
Jul 28, 2017Z-0262-2018Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.Terminated

Other KNS Clearances

The newest 510(k)s for the same product code; their median review was 118 days; see every KNS device

Record

Type
Special
Received
Feb 16, 2016
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.4300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files