Arca
Skip to content
Tools
Tools

Octavius I, Octavius Ii, Octavius Iii, Octavius 4D

Accelerator, Linear, Medical. 510(k) K160405, from Ptw-Freiburg Physikalisch-Technische-Werkstaetten (Freiburg, DE). Product code IYE, class 2.

As of Nov 18, 2026, K160405 is an FDA 510(k) held by Ptw-Freiburg Physikalisch-Technische-Werkstaetten, substantially equivalent on Oct 20, 2017 after 612 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 612 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The OCTAVIUS System is intended to collect beam data for patient plan verification of a treatment planning system (TPS) and under the aspect of machine QA for the following purposes: - IMRT patient plan verification - periodic QA procedures, e.g.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K160405, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 7, 2023Z-2577-2023Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the…Open
Aug 27, 2021Z-0479-2022When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Feb 16, 2016
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.