Giraffe Incubator Carestation CS1
Incubator, Neonatal. 510(k) K152809, from Ohmeda Medical, A Division of Datex-Ohmeda, Inc. (Laurel, MD). Product code FMZ, class 2.
As of Nov 18, 2026, K152809 is an FDA 510(k) held by Ohmeda Medical, A Division of Datex-Ohmeda, Inc., substantially equivalent on Dec 31, 2015 after 94 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 31, 2015
- Review time: 94 days
- Recalls: 3
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Giraffe Incubator Carestation is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Giraffe IncubatorK101778 · Ohmeda Medical · Jul 21, 2010
- Modification To: Ohmeda Medical Giraffe Incubator (Giraffe And Panda Uninterruptible Power Supply)K072512 · Datex-Ohmeda, Inc. · Oct 3, 2007
- Ohmeda Medical Giraffe IncubatorK020547 · Ohmeda Medical · May 9, 2002
- Ohmeda Medical Giraffe IncubatorK010222 · Ohmeda Medical · Mar 21, 2001
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Giraffe Omnibed Carestation CS1K213551 · Datex Ohmeda, Inc. · Mar 2, 2022
Recalls
Recalls FDA ties to K152809, newest first; each opens FDA's recall record| Jun 6, 2023 | Z-0141-2025Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week.… | Completed |
| Oct 11, 2019 | Z-0540-2020Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall. | Terminated |
| Sep 25, 2018 | Z-0480-2019Potential electric leakage affecting the user. | Terminated |
Other FMZ Clearances
The newest 510(k)s for the same product code; their median review was 179 days; see every FMZ device- Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)K260326 · Datex-Ohmeda, Inc. · Jul 28, 2026
- Babyleo TN500K251619 · Drägerwerk AG & Co. KGaA · Feb 13, 2026
- Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)K251663 · Datex-Ohmeda, Inc. · Oct 22, 2025
- mOm Essential Incubator (ME1)K243437 · Mom Incubators Limited · Aug 14, 2025
- Giraffe Incubator Carestation CS1K213553 · Datex Ohmeda, Inc. · Mar 2, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Sep 28, 2015
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files