Arca
Skip to content

Resection Electrodes

Electrode, Electrosurgical, Active, Urological. 510(k) K152092, from Olympus Winter & Ibe GmbH (Hamburg, DE). Product code FAS, class 2.

As of Nov 17, 2026, K152092 is an FDA 510(k) held by Olympus Winter & Ibe GmbH, substantially equivalent on Oct 26, 2015 after 90 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Oct 26, 2015
  • Review time: 90 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications. Cutting, ablation, resection, vaporization and coagulation with HF current.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K152092, newest first; each opens FDA's recall record
StartedReasonStatus
Jun 8, 2023Z-2413-2023Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,…Open, Classified

Other FAS Clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every FAS device

Record

Type
Traditional
Received
Jul 28, 2015
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.4300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files