Carestation 620/650/650C
Gas-Machine, Anesthesia. 510(k) K151570, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.
As of Nov 17, 2026, K151570 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Jan 21, 2016 after 225 days, with 14 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 21, 2016
- Review time: 225 days
- Recalls: 14
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Carestation 620/650/650c anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ge Datex-Ohmeda Avance CS2K131945 · Datex-Ohmeda · Aug 8, 2013
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- Carestation 750/750cK213867 · Datex-Ohmeda, Inc. · Mar 28, 2023
- A8, A9 Anesthesia SystemK201957 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Mar 26, 2021
Recalls
Recalls FDA ties to K151570, newest first; each opens FDA's recall record| Nov 14, 2025 | Z-0913-2026Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia… | Open, Classified |
| Nov 14, 2025 | Z-0915-2026Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia… | Open, Classified |
| Mar 21, 2025 | Z-1631-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Mar 21, 2025 | Z-1628-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Mar 21, 2025 | Z-1632-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Mar 21, 2025 | Z-1634-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Mar 21, 2025 | Z-1624-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Mar 21, 2025 | Z-1627-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in… | Open, Classified |
| Nov 25, 2019 | Z-0813-2020GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high… | Terminated |
| Jul 24, 2018 | Z-2954-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the… | Terminated |
| Jul 24, 2018 | Z-2956-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the… | Terminated |
| Jul 24, 2018 | Z-2955-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the… | Terminated |
| Oct 11, 2017 | Z-0579-2018Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems. | Open, Classified |
| May 2, 2017 | Z-2703-2017Carestation 620, 650 and 650c has Potential for Elevated FiCO2 and Unexpected System Malfunction | Terminated |
Other BSZ Clearances
The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device- A5 Anesthesia System (A5); A7 Anesthesia System (A7)K253330 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jun 25, 2026
- Atlan (A100); Atlan (A100 XL); Atlan(A350); Atlan (A350 XL)K252746 · Drägerwerk AG Co. Kgaa · May 21, 2026
- Codonics Safe Labeling System-Point of Care Station (PCS) (SLS-700i and related 610i, 620i, 630i series enabled to…K251352 · Codonics.Incorporated · Jul 11, 2025
- AtlanK230931 · Drägerwerk AG Co. Kgaa · Jul 23, 2023
- Carestation 750/750cK213867 · Datex-Ohmeda, Inc. · Mar 28, 2023
- A8, A9 Anesthesia SystemK201957 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Mar 26, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 10, 2015
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files