Carestation 620/650/650C

Gas-Machine, Anesthesia. 510(k) K151570, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.

As of Nov 17, 2026, K151570 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Jan 21, 2016 after 225 days, with 14 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 21, 2016
  • Review time: 225 days
  • Recalls: 14
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Carestation 620/650/650c anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K151570, newest first; each opens FDA's recall record
Nov 14, 2025Z-0913-2026Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia…Open, Classified
Nov 14, 2025Z-0915-2026Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia…Open, Classified
Mar 21, 2025Z-1631-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Mar 21, 2025Z-1628-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Mar 21, 2025Z-1632-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Mar 21, 2025Z-1634-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Mar 21, 2025Z-1624-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Mar 21, 2025Z-1627-2025Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…Open, Classified
Nov 25, 2019Z-0813-2020GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high…Terminated
Jul 24, 2018Z-2954-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the…Terminated
Jul 24, 2018Z-2956-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the…Terminated
Jul 24, 2018Z-2955-2018On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the…Terminated
Oct 11, 2017Z-0579-2018Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.Open, Classified
May 2, 2017Z-2703-2017Carestation 620, 650 and 650c has Potential for Elevated FiCO2 and Unexpected System MalfunctionTerminated

Other BSZ Clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Traditional
Received
Jun 10, 2015
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files