Arca
Skip to content
Tools
Tools

Oscar 2

System, Measurement, Blood-Pressure, Non-Invasive. 510(k) K151520, from Suntech Medical, Inc. (Morrisville, NC). Product code DXN, class 2.

As of Nov 18, 2026, K151520 is an FDA 510(k) held by Suntech Medical, Inc., substantially equivalent on Feb 19, 2016 after 259 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 19, 2016
  • Review time: 259 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Oscar 2, Model 250 system is a non-invasive oscillometric ambulatory blood pressure monitor that is intended to be used with AccuWin Pro, a PC-based computer program for the recording and displaying of up to 250 measurements of systolic and diastolic blood pressure and heart rate.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K151520, newest first; each opens FDA's recall record
Apr 20, 2020Z-2429-2020The device display, when activated, can show an incorrect decimal point when the blood pressure reading is a three-digit number and the unit of measure is set for mmHg.Terminated
Dec 11, 2019Z-0810-2020The Bravo Mini (Model 250D), Agilis Mini (Model 250D), and NBP One (Model 250D) are being recalled because of a user safety issue when rechargeable batteries are misused in the product.Terminated

Other DXN Clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every DXN device

Record

Type
Abbreviated
Received
Jun 5, 2015
Review panel
Cardiovascular
Regulation
21 CFR 870.1130

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.