SenTec Digital Monitor, OxiVenT Sensor, V-Sign Sensor, Staysite Adhesive Pad, V-STATS PC Software including V-CareNet
Monitor, Carbon-Dioxide, Cutaneous. 510(k) K151329, from Sentec AG (Therwil, CH). Product code LKD, class 2.
As of Nov 20, 2026, K151329 is an FDA 510(k) held by Sentec AG, substantially equivalent on Dec 17, 2015 after 213 days, with 9 recalls tied to it in FDA's recall database.
- Review time: 213 days
- Recalls: 9
- Cited as a predicate by: 1 device
The SenTec Digital Monitoring System ā consisting of the SenTec Digital Monitor (SDM), Sensors and Accessories ā is indicated for continuous, noninvasive patient monitoring. The SenTec Digital Monitoring System is indicated for use in clinical and non-clinical settings such as hospitals, hospital-type facilities, intra-hospital transport environments, clinics, physician offices, ambulatoryā¦
- Sentec Digital Monitor, V-Sign Sensor and V-Sign Sensor 2 With Digital Sensor Adaper CableK101690 Ā· Sentec AG Ā· Dec 3, 2010
- Tcm Tosca Monitoring System, Tcm Combim Monitoring SystemK093154 Ā· Radiometer Medical Aps Ā· Feb 5, 2010
- Sentec Digital Monitoring System (SDMS) tCOM+K232516 Ā· Sentec AG Ā· May 17, 2024
| Jul 25, 2023 | Z-2592-2023During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance. | Open |
| Aug 22, 2022 | Z-0148-2023Weak skin adhesion. | Terminated |
| Dec 10, 2020 | Z-1134-2021Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards. | Terminated |
| May 2, 2019 | Z-2153-2019, Z-2154-2019, Z-2155-2019, Z-2156-2019, Z-2152-2019There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure. | Terminated |
| Sep 15, 2016 | Z-0088-2017Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the⦠| Terminated |
- Sentec Digital Monitoring System (SDMS) tCOM+K232516 Ā· Sentec AG Ā· May 17, 2024
- Document
- 510(k) summary (PDF) ā
- Type
- Traditional
- Received
- May 18, 2015
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.2480
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files