Arca
Skip to content
Tools
Tools

Link® Megasystem-C®

Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer. 510(k) K151008, from Waldemar Link GmbH & Co. KG (Hamburg, DE). Product code KRO, class 2.

As of Nov 20, 2026, K151008 is an FDA 510(k) held by Waldemar Link GmbH & Co. KG, substantially equivalent on Aug 26, 2015 after 133 days, with 16 recalls tied to it in FDA's recall database.

  • Review time: 133 days
  • Recalls: 16
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The LINK® MEGASYSTEM-C® is intended to be used with the components of the Endo-Model® SL® Rotating and Non-Rotating Hinge Knee or Endo-Model® Knee System (#K143179) which can be integrated for knee joint replacement and with the MP® Reconstruction Prosthesis (#K142187) for hip replacement.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K151008, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 12, 2026Z-1511-2026, Z-1520-2026, Z-1512-2026, Z-1513-2026, Z-1516-2026, Z-1517-2026, Z-1523-2026, Z-1518-2026, Z-1514-2026, Z-1522-2026, Z-1521-2026, Z-1515-2026, Z-1519-2026The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.Open
Nov 10, 2025Z-0890-2026The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.Open
May 22, 2024Z-2411-2024, Z-2410-2024Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing techniqueOpen

Other KRO clearances

The newest 510(k)s for the same product code; their median review was 87 days; see every KRO device

Record

Type
Traditional
Received
Apr 15, 2015
Review panel
Orthopedic
Regulation
21 CFR 888.3510

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files