Calculator, Drug Dose. 510(k) K150910, from Roche Diabetes Care, Inc. (Indianapolis, IN). Product code NDC, class 2.

As of Nov 19, 2026, K150910 is an FDA 510(k) held by Roche Diabetes Care, Inc., substantially equivalent on Jun 3, 2015 after 61 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 61 days
  • Recalls: 6
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ACCU-CHEK Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K150910, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 3, 2019Z-2753-2019Users with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.Terminated
Jun 8, 2017Z-0625-2018Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may…Terminated
Jun 8, 2017Z-0197-2018Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown…Terminated
Mar 14, 2017Z-1899-2017A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.Terminated
Dec 20, 2016Z-1099-2017iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and…Terminated
Sep 14, 2016Z-0586-2017Roche Diabetes Care, Inc. initiated a voluntary recall for the version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS), released on July 11, 2016, due to the software containing a programming error (bug) in the…Terminated

Other NDC clearances

The newest 510(k)s for the same product code; their median review was 85 days; see every NDC device

Record

Type
Special
Received
Apr 3, 2015
Review panel
Anesthesiology
Regulation
21 CFR 868.1890

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files