Arca
Skip to content

Response 5.5/6.0 Spine System

Thoracolumbosacral Pedicle Screw System. 510(k) K150600, from OrthoPediatrics Corp. (Warsaw, IL). Product code NKB, class 2.

As of Nov 17, 2026, K150600 is an FDA 510(k) held by OrthoPediatrics Corp., substantially equivalent on May 22, 2015 after 73 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 22, 2015
  • Review time: 73 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K150600, newest first; each opens FDA's recall record
Jul 6, 2020Z-2742-2020Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the deviceOpen, Classified
Aug 3, 2017Z-0597-2018The 7mm diameter uniaxial pedicle screws have been color anodized with the same color utilized on 5 mm diameter screws.Terminated
Nov 10, 2015Z-0622-2016Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.Terminated

Other NKB Clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every NKB device

Record

Type
Traditional
Received
Mar 10, 2015
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.