Arca
Skip to content
Tools
Tools

SCENARIA Phase 3 Whole-body X-ray CT System

System, X-Ray, Tomography, Computed. 510(k) K150595, from Hitachi Medical Systems America, Inc. (Twinsburg, OH). Product code JAK, class 2.

As of Nov 26, 2026, K150595 is an FDA 510(k) held by Hitachi Medical Systems America, Inc., substantially equivalent on Oct 30, 2015 after 235 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 235 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SCENARIO Phase 3 system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the SCENARIO Phase 3 can be post processed by the SCENARIO Phase 3 to provide additional information.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • SCENARIA ViewK200498 · Hitachi Healthcare Americas · Jul 17, 2020
  • SCENARIA ViewK190841 · Hitachi Healthcare Americas · Sep 13, 2019

Recalls

Recalls FDA ties to K150595, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 16, 2019Z-0271-2020There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.Terminated
Jun 29, 2018Z-2839-2018The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The…Terminated

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Mar 9, 2015
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.