Panorama Patient Monitoring Network
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K150082, from Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Shenzhen, CN). Product code MHX, class 2.
As of Nov 18, 2026, K150082 is an FDA 510(k) held by Shenzhen Mindray Bio-Medical Electronics Co., Ltd., substantially equivalent on Sep 2, 2015 after 230 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Sep 2, 2015
- Review time: 230 days
- Recalls: 1
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe indications for use for the Panorama Patient Monitoring Network include: • Viewing real time patient clinical and demographic data • Graphical and numeric trending of clinical data • Storing and printing of clinical and demographic data • Setting independent alarm limits for data sent by the bedside monitor The clinical data displayed by the Panorama Patient Monitoring Network is obtained…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Panorama Patient Monitoring NetworkK142601 · Shenzhen Mindray Bio-Medical Electronics Co. · Nov 14, 2014
- Hypervisor Central Monitoring System (Including Telemetry Monitoring System, TMS-6016)K132036 · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · Jul 1, 2014
- St/Ar St And Arrhythmia Software Model: Release K.OK101521 · Philips Medical Systems · Jul 29, 2010
Recalls
Recalls FDA ties to K150082, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Dec 15, 2015 | Z-0701-2016An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings… | Terminated |
Other MHX Clearances
The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- M12 Telemetry SystemK252840 · Global Instrumentation, LLC · May 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 15, 2015
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files