Arca
Skip to content
Tools
Tools

Panorama Patient Monitoring Network

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K150082, from Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Shenzhen, CN). Product code MHX, class 2.

As of Nov 18, 2026, K150082 is an FDA 510(k) held by Shenzhen Mindray Bio-Medical Electronics Co., Ltd., substantially equivalent on Sep 2, 2015 after 230 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Sep 2, 2015
  • Review time: 230 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The indications for use for the Panorama Patient Monitoring Network include: • Viewing real time patient clinical and demographic data • Graphical and numeric trending of clinical data • Storing and printing of clinical and demographic data • Setting independent alarm limits for data sent by the bedside monitor The clinical data displayed by the Panorama Patient Monitoring Network is obtained…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K150082, newest first; each opens FDA's recall record
StartedReasonStatus
Dec 15, 2015Z-0701-2016An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Traditional
Received
Jan 15, 2015
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files