Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)

Respiratory Virus Panel Nucleic Acid Assay System. 510(k) K143653, from Nanosphere, Inc. (Northbrook, IL). Product code OCC, class 2.

As of Nov 20, 2026, K143653 is an FDA 510(k) held by Nanosphere, Inc., substantially equivalent on Sep 4, 2015 after 255 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 255 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K143653, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 5, 2022Z-1092-2022Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.Open
Oct 30, 2020Z-2006-2021Potential for a false-negative result.Open
Oct 13, 2020Z-0477-2021, Z-0476-2021Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using…Terminated
Dec 10, 2018Z-0054-2021No results/incorrect results due to failure of the hybridization heater.Open

Other OCC clearances

The newest 510(k)s for the same product code; their median review was 29 days; see every OCC device

Record

Type
Traditional
Received
Dec 23, 2014
Review panel
Microbiology
Regulation
21 CFR 866.3980

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files