Intervertebral Fusion Device With Integrated Fixation, Cervical. 510(k) K142940, from Titan Spine, LLC (Mequon, WI). Product code OVE, class 2.

As of Nov 17, 2026, K142940 is an FDA 510(k) held by Titan Spine, LLC, substantially equivalent on Jan 28, 2015 after 111 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 28, 2015
  • Review time: 111 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Endoskeleton® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K142940, newest first; each opens FDA's recall record
Jul 20, 2015Z-2327-2015Titan Spine is recalling specific Endoskeleton TCS Integrated Locking Bone Screw Assembly due to the locking bone screw collars not 'locking'.Terminated
Feb 20, 2015Z-1324-2015The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight…Terminated

Other OVE Clearances

The newest 510(k)s for the same product code; their median review was 66 days; see every OVE device

Record

Type
Traditional
Received
Oct 9, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.