Ziehm Vision RFD 3D
Interventional Fluoroscopic X-Ray System. 510(k) K142740, from Ziehm Imaging GmbH (Nuremberg, DE). Product code OWB, class 2.
As of Nov 17, 2026, K142740 is an FDA 510(k) held by Ziehm Imaging GmbH, substantially equivalent on Apr 6, 2015 after 194 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Apr 6, 2015
- Review time: 194 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ziehm Vision RfdK132904 · Ziehm Imaging GmbH · Dec 5, 2013
- Ziehm Vision 2K073346 · Ziehm Imaging, Inc. · Mar 21, 2008
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Mobile ImagingRing System, IRm, Loop-X Mobile Imaging Robot, Loop-XK203281 · Medphoton GmbH · Dec 22, 2020
- Ziehm Vision RFD 3DK202360 · Ziehm Imaging GmbH · Sep 16, 2020
- nView system 1K190064 · Nview Medical, Inc. · Jul 8, 2019
Recalls
Recalls FDA ties to K142740, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Apr 15, 2016 | Z-2639-2016Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted. | Terminated |
Other OWB Clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 24, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files