Biomet Headless Compression and Twist-off Screws
Screw, Fixation, Bone. 510(k) K142658, from Biomet Manufacturing Corp (Warsaw, IN). Product code HWC, class 2.
As of Nov 17, 2026, K142658 is an FDA 510(k) held by Biomet Manufacturing Corp, substantially equivalent on Nov 13, 2014 after 56 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 13, 2014
- Review time: 56 days
- Recalls: 2
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Biomet Headless Compression Screws and Twist-Off Screws are indicated for fixation of bone fractures, fusion of a joint (arthrodesis) or bone reconstruction (osteotomy) of the mid-foot bones, metatarsal and phalanges of the foot or the phalanges, metacarpals and carpals of the hand.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Frs ScrewK062352 · DePuy Orthopaedics, Inc. · Oct 19, 2006
- Twist-Off (Tm) ScrewK971069 · Landos, Inc. · Jul 9, 1997
- Twist-Off ScrewK962233 · Landos, Inc. · Oct 3, 1996
- Darco Small ScrewK082320 · Wrightmedicaltechnologyinc · Nov 5, 2008
- Memometal Fixos ScrewsK070039 · Memometal Technologies · Mar 21, 2007
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Monster® Screw SystemK190586 · Paragon 28, Inc. · Apr 3, 2019
- Biomet Headless Compression and Twist-Off ScrewsK190035 · Biomet, Inc. · Feb 7, 2019
- DePuy Synthes 2.0 mm Quick Insertion ScrewsK180541 · Synthes USA Products, LLC · Nov 21, 2018
Recalls
Recalls FDA ties to K142658, newest first; each opens FDA's recall record| Mar 28, 2019 | Z-1302-2019There is a potential of fracture, bending or shearing of the driver. | Terminated |
| Mar 28, 2019 | Z-1303-2019There is a potential of fracture, bending or shearing of the driver. | Terminated |
Other HWC Clearances
The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device- Ntegrate Threaded Nail SystemK261961 · Nvision Biomedical Technologies · Oct 14, 2026
- OsteoCentric Technologies Cannulated and Soild Fasteners with NutsK263116 · OsteoCentric Technologies · Oct 2, 2026
- Arthrex Beaming SystemK262975 · Arthrex, Inc. · Sep 25, 2026
- DynaNail Helix® Fixation SystemK260995 · MedShape, Inc. · Sep 23, 2026
- F.A.S.T ScrewK254280 · Steps Ortho · Sep 21, 2026
- ActivaScrew Cannulated COMK261057 · Bioretec, Ltd. · Sep 4, 2026
- OsteoCoil GT Active Compression SystemK260106 · Flower Orthopedics Corporation DBA Conventus Flower Orthoped · Aug 24, 2026
- Large Compression ScrewsK262344 · TriMed, Inc. · Aug 12, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 18, 2014
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files