Arca
Skip to content

Chesapeake Stabilization System

Intervertebral Fusion Device With Integrated Fixation, Lumbar. 510(k) K142487, from K2m, Inc. (Leesburg, VA). Product code OVD, class 2.

As of Nov 17, 2026, K142487 is an FDA 510(k) held by K2m, Inc., substantially equivalent on May 7, 2015 after 245 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: May 7, 2015
  • Review time: 245 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

When used as a cervical intervertebral body fusion device, the CHESAPEAKE Stabilization System implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K142487, newest first; each opens FDA's recall record
StartedReasonStatus
Sep 8, 2021Z-0104-2022Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery.Terminated

Other OVD Clearances

The newest 510(k)s for the same product code; their median review was 95 days; see every OVD device

Record

Type
Traditional
Received
Sep 4, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.