Chesapeake Stabilization System
Intervertebral Fusion Device With Integrated Fixation, Lumbar. 510(k) K142487, from K2m, Inc. (Leesburg, VA). Product code OVD, class 2.
As of Nov 17, 2026, K142487 is an FDA 510(k) held by K2m, Inc., substantially equivalent on May 7, 2015 after 245 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: May 7, 2015
- Review time: 245 days
- Recalls: 1
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFWhen used as a cervical intervertebral body fusion device, the CHESAPEAKE Stabilization System implants are indicated for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Chesapeake Spinal SystemK092211 · K2m, Inc. · Mar 5, 2010
- Chesapeake (Cervical Components)K111439 · K2m, Inc. · Aug 24, 2011
- Chesapeake Spinal SystemK133494 · K2m, Inc. · Jun 10, 2014
- Aleutian Ibf SystemK082698 · K2m, Inc. · Apr 21, 2009
- Aleutian Ibf SystemK133614 · K2m, Inc. · Apr 3, 2014
- Brigade Standalone SystemK123045 · Nuvasive, Inc. · Apr 16, 2013
Recalls
Recalls FDA ties to K142487, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Sep 8, 2021 | Z-0104-2022Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery. | Terminated |
Other OVD Clearances
The newest 510(k)s for the same product code; their median review was 95 days; see every OVD device- Innovasis AxTiHA® Stand-Alone ALIF System;Innovasis TxTiHA® IBF System;Innovasis…K262025 · INNO Holdings Inc. DBA Innovasis · Sep 11, 2026
- SABER-XA SystemK261408 · Elevation Spine · Aug 3, 2026
- Interwedge® Standalone LateralK261809 · Foundation Surgical Group · Jul 31, 2026
- Tesera-K Anterior Lumbar Interbody Fusion (ALIF) System; Tesera-K XL System; Lumbar…K261353 · Kyocera Medical Technologies, Inc. · Jul 15, 2026
- Anteralign LS coverplateK260810 · Tyber Medical, LLC · Jun 10, 2026
- MectaLIF 3D Metal AnteriorK254202 · Medacta International S.A. · May 5, 2026
- SCRIPT Implant SystemK253401 · Globus Medical, Inc. · Apr 29, 2026
- Ventana A Anterior Lumbar Interbody SystemK253559 · Spinal Elements, Inc. · Feb 25, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 4, 2014
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3080
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files