Arca
Skip to content
Tools
Tools

NS3000 Holmium Laser System

Powered Laser Surgical Instrument. 510(k) K142286, from New Star Lasers, Inc. (Roseville, CA). Product code GEX, class 2.

As of Nov 19, 2026, K142286 is an FDA 510(k) held by New Star Lasers, Inc., substantially equivalent on Oct 31, 2014 after 74 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 74 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The NS3000 laser system is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K142286, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
May 13, 2015Z-1918-2015During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value…Terminated

Other GEX clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every GEX device

Record

Type
Special
Received
Aug 18, 2014
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4810

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.