Symbia 6.0
System, Tomography, Computed, Emission. 510(k) K142006, from Siemens Medical Solutions USA, Inc. (Hoffman Estates, IL). Product code KPS, class 2.
As of Nov 17, 2026, K142006 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Aug 21, 2014 after 29 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 21, 2014
- Review time: 29 days
- Recalls: 9
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- SymbiaK131634 · Siemens Medical Solutions USA, Inc. · Aug 16, 2013
- Somatom Emotion 6 Ct Systems, Somatom Emotion 16K133424 · Siemens Medical Solutions USA, Inc. · Jan 17, 2014
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Symbia Intevo BoldK162483 · Siemens Medical Solutions USA, Inc. · Jan 9, 2017
- Symbia 6.5K162337 · Siemens Medical Solutions USA, Inc. · Sep 8, 2016
- syngo.via MI WorkflowsK160426 · Siemens Medical Solutions USA, Inc. · Mar 4, 2016
Recalls
Recalls FDA ties to K142006, newest first; each opens FDA's recall record| Aug 5, 2016 | Z-2758-2016"We have received reports of patients hair being caught in the e.cam and Symbia E patient bed up/down drive. The purpose of this letter is to remind users of the patient positioning instructions for your e.cam or Symbia… | Terminated |
| Mar 31, 2016 | Z-2898-2020E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote. SPECT: To… | Terminated |
| Feb 26, 2016 | Z-1262-2016We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that… | Terminated |
| Feb 26, 2016 | Z-1263-2016We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that… | Terminated |
| Dec 31, 2015 | Z-0687-2016To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not sent in its entirety when the e.cam systems were refurbished and shipped to the customer. | Terminated |
| Jul 31, 2015 | Z-2436-2015Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to… | Terminated |
| Jul 31, 2015 | Z-2437-2015Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to… | Terminated |
| Jul 31, 2015 | Z-2435-2015Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to… | Terminated |
| Jun 26, 2015 | Z-0208-2016Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to… | Terminated |
Other KPS Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jul 23, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files