Arca
Skip to content
Tools
Tools

Nakoma Acp

Appliance, Fixation, Spinal Intervertebral Body. 510(k) K141993, from Alliance Partners, LLC (Memphis, TN). Product code KWQ, class 2.

As of Nov 18, 2026, K141993 is an FDA 510(k) held by Alliance Partners, LLC, substantially equivalent on Nov 6, 2014 after 107 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Nov 6, 2014
  • Review time: 107 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Nakoma ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patents with the following indications: • Degenerative Disk Disease (as defined by neck pain of discogenic origin with degeneration of the disc…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K141993, newest first; each opens FDA's recall record
StartedReasonStatus
Jun 9, 2015Z-1989-2015Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.Terminated

Other KWQ Clearances

The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device

Record

Type
Traditional
Received
Jul 22, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3060

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.