Nakoma Acp
Appliance, Fixation, Spinal Intervertebral Body. 510(k) K141993, from Alliance Partners, LLC (Memphis, TN). Product code KWQ, class 2.
As of Nov 18, 2026, K141993 is an FDA 510(k) held by Alliance Partners, LLC, substantially equivalent on Nov 6, 2014 after 107 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Nov 6, 2014
- Review time: 107 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Nakoma ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patents with the following indications: • Degenerative Disk Disease (as defined by neck pain of discogenic origin with degeneration of the disc…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Custom Spine Regent Anterior Cervical PlateK091134 · Custom Spine, Inc. · Aug 17, 2009
- Quintex Cervical Plating SystemK100243 · Aesculap Implant Systems, LLC · Sep 2, 2010
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Skyway Anterior Cervical Plate SystemK231766 · Kyocera Medical Technologies Inc. (KMTI) · Sep 12, 2023
- Medina Anterior Cervical Plate SystemK201521 · Alliance Partners, LLC · Aug 6, 2020
- Alamo® CK173128 · Alliance Partners, LLC · Mar 22, 2018
Recalls
Recalls FDA ties to K141993, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jun 9, 2015 | Z-1989-2015Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation. | Terminated |
Other KWQ Clearances
The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device- corra® cervical plating systemK262149 · Carlsmed, Inc. · Sep 24, 2026
- Palisades Buttress Plate SystemK261309 · Bethesda Medical Devices, LLC · Jul 16, 2026
- ARIES Anterior Cervical Plate SystemsK252776 · ZheJiang Decans Medical Devices Co., Ltd. · May 14, 2026
- CastleLoc-P Anterior Cervical Plate SystemK261112 · L&K BIOMED Co., Ltd. · May 6, 2026
- corra cervical plating systemK260570 · Carlsmed, Inc. · Apr 22, 2026
- VyPlate Anterior Cervical Plate SystemK260697 · Vy Spine, LLC · Mar 27, 2026
- ANTERIS Thoracolumbar Plate SystemK260015 · SpineCraft · Mar 4, 2026
- PathLoc Lumbar Plate SystemK251940 · L&K BIOMED Co., Ltd. · Mar 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 22, 2014
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3060
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files