Arca
Skip to content

Zimmer Periarticular Locking System

Plate, Fixation, Bone. 510(k) K141734, from Zimmer, Inc. (Warsaw, IN). Product code HRS, class 2.

As of Nov 17, 2026, K141734 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Aug 13, 2014 after 47 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 13, 2014
  • Review time: 47 days
  • Recalls: 3

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Zimmer Periarticular Locking Plate System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including: - Comminuted fractures - Supracondylar fractures - Intra-articular and extra-articular condylar fractures - Fractures in osteopenic bone - Nonunions - Malunions

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K141734, newest first; each opens FDA's recall record
Jun 27, 2023Z-2254-2023Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.Open, Classified
Oct 10, 2019Z-0317-2020Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).Terminated
Oct 10, 2019Z-0310-2020Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).Terminated

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
Jun 27, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.