Arca
Skip to content
Tools
Tools

Sysmex XN Series

Counter, Differential Cell. 510(k) K141681, from Sysmex America, Inc. (Lincolnshire, IL). Product code GKZ, class 2.

As of Nov 19, 2026, K141681 is an FDA 510(k) held by Sysmex America, Inc., substantially equivalent on Dec 5, 2014 after 165 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 165 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The XN-Series modules (XN-11, XN-21) are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical and reference laboratories.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K141681, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Mar 23, 2017Z-2034-2017Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on…Terminated

Other GKZ clearances

The newest 510(k)s for the same product code; their median review was 188 days; see every GKZ device

Record

Type
Traditional
Received
Jun 23, 2014
Review panel
Hematology
Regulation
21 CFR 864.5220

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files