CP150 Electrocardiograph With Spirometry Option
Electrocardiograph. 510(k) K141582, from Welch Allyn (Skaneateles Falls, NY). Product code DPS, class 2.
As of Nov 19, 2026, K141582 is an FDA 510(k) held by Welch Allyn, substantially equivalent on Jul 11, 2014 after 28 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 28 days
- Recalls: 2
The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and measure patient cardiac function. The 12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
- CP150 ElectrocardiographK131573 · Welch Allyn, Inc. · Jul 3, 2013
- Medikro Nano, Medikro Primo and Medikro ProK133428 · Medikro OY · May 21, 2014
| Mar 24, 2025 | Z-1560-2025There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter. | Open |
| Jun 3, 2024 | Z-2323-2024During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity… | Open |
- Kinetic Software LibraryK260640 · Monebo Technologies, Inc. · Oct 23, 2026
- MyoVista wavECG (DG-200-B15-003)K260005 · HeartSciences, Inc. · Oct 5, 2026
- QCA 10 Electrocardiograph (QCARD-10); QCA 10W Electrocardiograph (QCARD-10W)K260250 · QT Medical, Inc. · Sep 29, 2026
- SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)K252463 · Edan Instruments, Inc. · Jul 2, 2026
- MAC 7 Resting ECG Analysis SystemK251670 · Ge Medical Systems Information Technologies, Inc. · Jan 26, 2026
- AccurECG Analysis System (v2.0)K252361 · Accurkardia, Inc. · Dec 22, 2025
- Withings BeamO (SCT02)K252474 · Withings · Nov 10, 2025
- Cardiologs Holter PlatformK250569 · Philips France Commercial · Aug 6, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jun 13, 2014
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2340
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files