Expel Apd Drainage Catheter System

Stents, Drains And Dilators For The Biliary Ducts. 510(k) K141335, from Boston Scientific Corp (Maple Grove, MN). Product code FGE, class 2.

As of Nov 17, 2026, K141335 is an FDA 510(k) held by Boston Scientific Corp, substantially equivalent on Oct 17, 2014 after 149 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 17, 2014
  • Review time: 149 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The drainage catheter is intended to provide percutaneous drainage of abscess fluid and biliary collections. 510(K) Number (f / known) K 141335 This section applies only to requirements of the Paperwork Reduction Act of 1995.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K141335, newest first; each opens FDA's recall record
Feb 25, 2015Z-1355-2015Complaints of device fragmentation after the catheter was implanted in the biliary system. The most serious patient risk for this issue is additional intervention for fragment retrieval using minimally invasive methods…Terminated
Feb 25, 2015Z-1336-2015Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD…Terminated

Other FGE Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every FGE device

Record

Type
Traditional
Received
May 21, 2014
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5010

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files