Acculif Tl And Pl Cage
Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K141217, from Stryker Corporation (Allendale, NJ). Product code MAX, class 2.
As of Nov 18, 2026, K141217 is an FDA 510(k) held by Stryker Corporation, substantially equivalent on Jul 16, 2014 after 65 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 16, 2014
- Review time: 65 days
- Recalls: 3
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIntervertebral Body Fusion Device: The AccuLIF TL and PL Cages are indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one level or two continuous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Acculif Tl And Pl CageK132505 · Coalign Innovations, Inc. · Sep 12, 2013
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Roccia Alif, Roccia TlifK180963 · Silony Medical GmbH · Jul 11, 2018
- Cortina™ Lumbar Cage SystemK171914 · Neurostructures, Inc. · Oct 25, 2017
- Omega LIF SystemK150395 · Amendia, Inc. · Aug 10, 2015
- AVS® AL and ALign PEEK Spacers, AVS® PL and UniLIF PEEK Spacers, AVS® TL PEEK Spacer, AVS® Navigator PEEK Spacer, AVS®…K143163 · Stryker Corporation · Jan 26, 2015
Recalls
Recalls FDA ties to K141217, newest first; each opens FDA's recall record| Jul 26, 2016 | Z-2788-2016Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported… | Terminated |
| Jul 26, 2016 | Z-2787-2016Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported… | Terminated |
| Jul 26, 2016 | Z-2786-2016Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported… | Terminated |
Other MAX Clearances
The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device- Symmetrix PX Spinal SystemK262572 · Medtronic Sofamor Danek USA, Inc. · Oct 29, 2026
- Calibrate MAX System; Calibrate NanoTec MAX System; Calibrate CCX Interbody System; Calibrate NanoTec CCX Interbody…K262646 · Alphatec Spine, Inc. · Oct 21, 2026
- Scala® TlK261049 · Spineart SA · Aug 20, 2026
- TruLift Expandable SystemK261430 · Life Spine, Inc. · Jul 28, 2026
- Catalyft PL Expandable Interbody SystemK253011 · Medtronic Sofamor Danek USA, Inc. · Jul 17, 2026
- Symmetrix L Spinal SystemK261020 · Medtronic Sofamor Danek USA, Inc. · Jul 14, 2026
- Admiral ACP System; Explorer TO System; Laminoplasty System; Manta Ray TDF Spacer; Meridian Interbody System; Meridian…K253420 · SeaSpine Orthopedics Corporation · Jul 1, 2026
- Fule Interbody Fusion Cage SystemK253151 · Beijing Fule Science & Technology Development Co., Ltd. · Jun 15, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 12, 2014
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3080
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files