Arca
Skip to content
Tools
Tools

Spacelabs Healthcare Telemetry Receiver

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K141156, from Spacelabs Healthcare (Nederland, CO). Product code MHX, class 2.

As of Nov 18, 2026, K141156 is an FDA 510(k) held by Spacelabs Healthcare, substantially equivalent on Sep 9, 2014 after 127 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 9, 2014
  • Review time: 127 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Spacelabs Healthcare Telemetry Receiver, Model 96280, is intended to provide the Spacelabs Healthcare monitoring system with adult, pediatric and neonatal patient data of patients connected to Spacelabs Healthcare telemetry transmitters.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K141156, newest first; each opens FDA's recall record
Nov 25, 2024Z-0810-2025Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…Open, Classified
Nov 25, 2024Z-0809-2025Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…Open, Classified
May 31, 2022Z-1513-2022Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal…Open, Classified
Jan 17, 2018Z-0532-2018Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.Terminated
Aug 25, 2016Z-2886-2016The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian…Terminated
Mar 28, 2016Z-1502-2016Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Traditional
Received
May 5, 2014
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files