Evotech Ecr
Accessories, Cleaning, For Endoscope. 510(k) K140977, from Advanced Sterilization Products (Irvin, CA). Product code FEB, class 2.
As of Nov 23, 2026, K140977 is an FDA 510(k) held by Advanced Sterilization Products, substantially equivalent on Dec 22, 2014 after 250 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 250 days
- Recalls: 3
- Cited as a predicate by: 1 device
The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX® OPA Concentrate and an enzymatic detergent (CIDEZYME GI) to achieve cleaning and high level disinfection of heat sensitive (>60 0C) semi-critical endoscopes.
- Evotech Endoscope Cleaner and ReprocessorK082392 · Advanced Sterilization Products · Nov 13, 2008
- Cidex Opa SolutionK991487 · Advanced Sterilization Products · Oct 8, 1999
- Biostar Oia ShigatoxK061889 · Inverness Medical-Biostar, Inc. · Dec 12, 2006
- Evotech EcrK152189 · Advanced Sterilization Products (Asp) · Jul 20, 2016
| Sep 27, 2016 | Z-0965-2017Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak. | Terminated |
| May 11, 2015 | Z-1714-2015, Z-1715-2015Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without… | Terminated |
- MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope ReprocessorK261791 · Steris · Oct 2, 2026
- System 83 Revolve Endoscope Washer/DisinfectorK241168 · Wassenburg Medical, Inc. · Dec 19, 2024
- enspire 3000 Cleaning and Liquid Chemical Sterilant Processing SystemK243433 · STERIS Corporation · Dec 2, 2024
- enspire 3000 Cleaning and Liquid Chemical Sterile Processing SystemK230930 · STERIS Corporation · Jun 30, 2023
- SCOPE BUDDY PLUS Endoscope Flushing AidK213833 · Medivators, Inc. · Jan 14, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 16, 2014
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files