Depuy Attune Knee System - Cementless Cr And Ps Femoral Componenets

Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer. 510(k) K140881, from Depuy(Ireland) (Co. Cork, IE). Product code MBH, class 2.

As of Nov 17, 2026, K140881 is an FDA 510(k) held by Depuy(Ireland), substantially equivalent on Feb 13, 2015 after 312 days, with 10 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 13, 2015
  • Review time: 312 days
  • Recalls: 10
  • Cited as a predicate by: 3 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K140881, newest first; each opens FDA's recall record
Jun 12, 2015Z-2140-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2137-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2139-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2138-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2145-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2143-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2142-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2146-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2141-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated
Jun 12, 2015Z-2144-2015Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal…Terminated

Other MBH Clearances

The newest 510(k)s for the same product code; their median review was 153 days; see every MBH device

Record

Type
Traditional
Received
Apr 7, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3565

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files