Dimension Vista Free Thyroxine Flex Reagent Cartridge, FT4, Dimension Vista Thyroid Stimulating Hormone Flex Reagent
Radioimmunoassay, Free Thyroxine. 510(k) K140859, from Siemens Healthcare Diagnostics, Inc. (Newark, DE). Product code CEC, class 2.
As of Nov 18, 2026, K140859 is an FDA 510(k) held by Siemens Healthcare Diagnostics, Inc., substantially equivalent on Nov 26, 2014 after 237 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 26, 2014
- Review time: 237 days
- Recalls: 3
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Dimension Vista Free Thyroxine Flex Reagent Cartridge And Loci 1 CalibratorK053531 · Dade Behring, Inc. · Mar 3, 2006
- Dimension Vista Thyroid Stimulating Hormone Flex Reagent CartridgeK060090 · Dade Behring, Inc. · Feb 27, 2006
Recalls
Recalls FDA ties to K140859, newest first; each opens FDA's recall record| Jan 2, 2018 | Z-1673-2018The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur… | Terminated |
| Mar 21, 2017 | Z-2611-2017The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH… | Terminated |
| Apr 20, 2015 | Z-1708-2015Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well. | Terminated |
Other CEC Clearances
The newest 510(k)s for the same product code; see every CEC device- Access Free T4K240273 · Beckman Coulter, Inc. · Mar 6, 2024
- Elecsys FT4 IVK220456 · Roche Diagnostics · Apr 7, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 3, 2014
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1695
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files